11 CALIFORNIA LEGISLATURE 20172018 REGULAR SESSION Assembly Bill No. 2419Introduced by Assembly Member FriedmanFebruary 14, 2018 An act to amend Section 130061 of the Health and Safety Code, relating to seismic safety. LEGISLATIVE COUNSEL'S DIGESTAB 2419, as introduced, Friedman. Seismic safety: hospitals.Existing law, the Alfred E. Alquist Hospital Facilities Seismic Safety Act of 1983, establishes, under the jurisdiction of the Office of Statewide Health Planning and Development, a program of seismic safety building standards for certain hospitals constructed on and after March 7, 1973.Existing law provides that, after January 1, 2008, a general acute care hospital building that is determined to be a potential risk of collapse or to pose significant loss of life in the event of seismic activity be used only for nonacute care hospital purposes, except that the office may grant a 5-year extension under prescribed circumstances. Existing law allows the office to grant a hospital that has received extensions under specified provisions an additional extension of up to 7 years for a hospital building that it owns or operates if the hospital meets specified milestones. Existing law requires an owner of a general acute care hospital building that is classified as nonconforming to submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the extension provisions, and to annually update the office with any changes or adjustments. Existing law requires the report to include the number of inpatient beds and patient days for the years 2008, 2009, and 2010. Existing law requires the office to make that information available on its Internet Web site, as specified. Existing law authorizes the imposition of fines for the failure to submit a report under those provisions.This bill would instead require that report to include the number of inpatient beds and patient days for the years 2008 to 2016, inclusive.Digest Key Vote: MAJORITY Appropriation: NO Fiscal Committee: YES Local Program: NO Bill TextThe people of the State of California do enact as follows:SECTION 1. Section 130061 of the Health and Safety Code is amended to read:130061. (a) An owner of a general acute care hospital building that is classified as a nonconforming Structural Performance Category-1 (SPC-1) building, who has not requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than April 15, 2007, describing the status of each building in complying with the requirements of Section 130060. The report shall identify at least all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of this action.(b) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, who has requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than June 30, 2009, describing the status of each building in complying with the requirements of Section 130060. The report shall identify, at a minimum, all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of that action.(c) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, shall submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the requirements of Section 130060, and annually thereafter shall update the office with any changes or adjustments. The report shall identify at least all of the following:(1) For each building that is subject to subdivision (a) of Section 130060 that is planned for retrofit or replacement, the report shall identify:(A) Whether the hospital owner intends to retrofit or replace the building to SPC-2, SPC-3, SPC-4, or SPC-5.(B) The deadline, as described in Section 130060 or 130061.5, for retrofit or replacement of the building that the hospital owner intends to meet, and the applicable extension for which the hospital owner has been approved.(C) The project number or numbers for retrofit or replacement of each building.(D) The projected construction start date or dates and projected construction completion date or dates.(E) The most recent project status and approvals.(F) The number of inpatient beds and patient days, by type of unit and type of service to be provided.(2) For the building or buildings to be removed from acute care service, the following information shall be included:(A) The projected date or dates the building will be removed from service.(B) The planned uses of the building or buildings to be removed from acute care service.(C) The inpatient services currently delivered in the building or buildings.(D) The number of inpatient beds and patient days, by type of unit and type of service, for the years 2008, 2009, and 2010. each of the years 2008 to 2016, inclusive.(E) Whether the general acute care services and beds will be relocated to a new or retrofitted building and any corresponding building sites or project numbers.(3) Each hospital owner shall also report, for each facility for which any buildings will be removed from acute care service, any net change in the number of inpatient beds, by type of unit and type of service, taking into account beds provided in buildings to be taken out of service, beds provided in buildings to be retrofitted or replaced, and beds provided in any other buildings used for general acute care inpatient services by the facility.(4) Each hospital owner shall report any general acute care hospital inpatient service that is provided in any general acute care hospital building that is rated SPC-1.(5) Each hospital owner shall report the final configuration of all buildings on its campus showing how each building will comply with the SPC-5/NPC-4 or 5 requirements, whether by retrofit or by replacement, and the type of services that will be provided in each general acute care hospital building.(d) The office shall make the information required by subdivisions (a) and (b), available on its Internet Web site within 90 days of receipt of this information.(e) The office shall make the information required by subdivision (c) available on its Internet Web site within 90 days of receipt of this information. The office shall include the hospital name, hospital owners, and location of the buildings included in the report, and, to the extent possible, for service areas containing buildings for which hospital owners report information pursuant to subdivision (c), include information on the number of inpatient beds, by type of unit and type of service, provided by facilities operating buildings that are classified as SPC-2, SPC-3, SPC-4, and SPC-5.(f) Hospitals that have not reported pursuant to this section are not eligible for the extension provided in subdivision (f) of Section 130060.(g) A hospital that has not submitted a report pursuant to this section shall be assessed a fine of ten dollars ($10) per licensed acute care bed per day, but in no case to exceed one thousand dollars ($1,000) per day for each SPC-1 building not in compliance with this section until it has complied with the provisions of this section. These fines shall be deposited into the Hospital Building Fund as specified pursuant to Section 129795. A hospital assessed a fine pursuant to this section may appeal the assessment to the Hospital Building Safety Board.
22
33 CALIFORNIA LEGISLATURE 20172018 REGULAR SESSION Assembly Bill No. 2419Introduced by Assembly Member FriedmanFebruary 14, 2018 An act to amend Section 130061 of the Health and Safety Code, relating to seismic safety. LEGISLATIVE COUNSEL'S DIGESTAB 2419, as introduced, Friedman. Seismic safety: hospitals.Existing law, the Alfred E. Alquist Hospital Facilities Seismic Safety Act of 1983, establishes, under the jurisdiction of the Office of Statewide Health Planning and Development, a program of seismic safety building standards for certain hospitals constructed on and after March 7, 1973.Existing law provides that, after January 1, 2008, a general acute care hospital building that is determined to be a potential risk of collapse or to pose significant loss of life in the event of seismic activity be used only for nonacute care hospital purposes, except that the office may grant a 5-year extension under prescribed circumstances. Existing law allows the office to grant a hospital that has received extensions under specified provisions an additional extension of up to 7 years for a hospital building that it owns or operates if the hospital meets specified milestones. Existing law requires an owner of a general acute care hospital building that is classified as nonconforming to submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the extension provisions, and to annually update the office with any changes or adjustments. Existing law requires the report to include the number of inpatient beds and patient days for the years 2008, 2009, and 2010. Existing law requires the office to make that information available on its Internet Web site, as specified. Existing law authorizes the imposition of fines for the failure to submit a report under those provisions.This bill would instead require that report to include the number of inpatient beds and patient days for the years 2008 to 2016, inclusive.Digest Key Vote: MAJORITY Appropriation: NO Fiscal Committee: YES Local Program: NO
44
55
66
77
88
99 CALIFORNIA LEGISLATURE 20172018 REGULAR SESSION
1010
1111 Assembly Bill No. 2419
1212
1313 Introduced by Assembly Member FriedmanFebruary 14, 2018
1414
1515 Introduced by Assembly Member Friedman
1616 February 14, 2018
1717
1818 An act to amend Section 130061 of the Health and Safety Code, relating to seismic safety.
1919
2020 LEGISLATIVE COUNSEL'S DIGEST
2121
2222 ## LEGISLATIVE COUNSEL'S DIGEST
2323
2424 AB 2419, as introduced, Friedman. Seismic safety: hospitals.
2525
2626 Existing law, the Alfred E. Alquist Hospital Facilities Seismic Safety Act of 1983, establishes, under the jurisdiction of the Office of Statewide Health Planning and Development, a program of seismic safety building standards for certain hospitals constructed on and after March 7, 1973.Existing law provides that, after January 1, 2008, a general acute care hospital building that is determined to be a potential risk of collapse or to pose significant loss of life in the event of seismic activity be used only for nonacute care hospital purposes, except that the office may grant a 5-year extension under prescribed circumstances. Existing law allows the office to grant a hospital that has received extensions under specified provisions an additional extension of up to 7 years for a hospital building that it owns or operates if the hospital meets specified milestones. Existing law requires an owner of a general acute care hospital building that is classified as nonconforming to submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the extension provisions, and to annually update the office with any changes or adjustments. Existing law requires the report to include the number of inpatient beds and patient days for the years 2008, 2009, and 2010. Existing law requires the office to make that information available on its Internet Web site, as specified. Existing law authorizes the imposition of fines for the failure to submit a report under those provisions.This bill would instead require that report to include the number of inpatient beds and patient days for the years 2008 to 2016, inclusive.
2727
2828 Existing law, the Alfred E. Alquist Hospital Facilities Seismic Safety Act of 1983, establishes, under the jurisdiction of the Office of Statewide Health Planning and Development, a program of seismic safety building standards for certain hospitals constructed on and after March 7, 1973.
2929
3030 Existing law provides that, after January 1, 2008, a general acute care hospital building that is determined to be a potential risk of collapse or to pose significant loss of life in the event of seismic activity be used only for nonacute care hospital purposes, except that the office may grant a 5-year extension under prescribed circumstances. Existing law allows the office to grant a hospital that has received extensions under specified provisions an additional extension of up to 7 years for a hospital building that it owns or operates if the hospital meets specified milestones. Existing law requires an owner of a general acute care hospital building that is classified as nonconforming to submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the extension provisions, and to annually update the office with any changes or adjustments. Existing law requires the report to include the number of inpatient beds and patient days for the years 2008, 2009, and 2010. Existing law requires the office to make that information available on its Internet Web site, as specified. Existing law authorizes the imposition of fines for the failure to submit a report under those provisions.
3131
3232 This bill would instead require that report to include the number of inpatient beds and patient days for the years 2008 to 2016, inclusive.
3333
3434 ## Digest Key
3535
3636 ## Bill Text
3737
3838 The people of the State of California do enact as follows:SECTION 1. Section 130061 of the Health and Safety Code is amended to read:130061. (a) An owner of a general acute care hospital building that is classified as a nonconforming Structural Performance Category-1 (SPC-1) building, who has not requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than April 15, 2007, describing the status of each building in complying with the requirements of Section 130060. The report shall identify at least all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of this action.(b) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, who has requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than June 30, 2009, describing the status of each building in complying with the requirements of Section 130060. The report shall identify, at a minimum, all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of that action.(c) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, shall submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the requirements of Section 130060, and annually thereafter shall update the office with any changes or adjustments. The report shall identify at least all of the following:(1) For each building that is subject to subdivision (a) of Section 130060 that is planned for retrofit or replacement, the report shall identify:(A) Whether the hospital owner intends to retrofit or replace the building to SPC-2, SPC-3, SPC-4, or SPC-5.(B) The deadline, as described in Section 130060 or 130061.5, for retrofit or replacement of the building that the hospital owner intends to meet, and the applicable extension for which the hospital owner has been approved.(C) The project number or numbers for retrofit or replacement of each building.(D) The projected construction start date or dates and projected construction completion date or dates.(E) The most recent project status and approvals.(F) The number of inpatient beds and patient days, by type of unit and type of service to be provided.(2) For the building or buildings to be removed from acute care service, the following information shall be included:(A) The projected date or dates the building will be removed from service.(B) The planned uses of the building or buildings to be removed from acute care service.(C) The inpatient services currently delivered in the building or buildings.(D) The number of inpatient beds and patient days, by type of unit and type of service, for the years 2008, 2009, and 2010. each of the years 2008 to 2016, inclusive.(E) Whether the general acute care services and beds will be relocated to a new or retrofitted building and any corresponding building sites or project numbers.(3) Each hospital owner shall also report, for each facility for which any buildings will be removed from acute care service, any net change in the number of inpatient beds, by type of unit and type of service, taking into account beds provided in buildings to be taken out of service, beds provided in buildings to be retrofitted or replaced, and beds provided in any other buildings used for general acute care inpatient services by the facility.(4) Each hospital owner shall report any general acute care hospital inpatient service that is provided in any general acute care hospital building that is rated SPC-1.(5) Each hospital owner shall report the final configuration of all buildings on its campus showing how each building will comply with the SPC-5/NPC-4 or 5 requirements, whether by retrofit or by replacement, and the type of services that will be provided in each general acute care hospital building.(d) The office shall make the information required by subdivisions (a) and (b), available on its Internet Web site within 90 days of receipt of this information.(e) The office shall make the information required by subdivision (c) available on its Internet Web site within 90 days of receipt of this information. The office shall include the hospital name, hospital owners, and location of the buildings included in the report, and, to the extent possible, for service areas containing buildings for which hospital owners report information pursuant to subdivision (c), include information on the number of inpatient beds, by type of unit and type of service, provided by facilities operating buildings that are classified as SPC-2, SPC-3, SPC-4, and SPC-5.(f) Hospitals that have not reported pursuant to this section are not eligible for the extension provided in subdivision (f) of Section 130060.(g) A hospital that has not submitted a report pursuant to this section shall be assessed a fine of ten dollars ($10) per licensed acute care bed per day, but in no case to exceed one thousand dollars ($1,000) per day for each SPC-1 building not in compliance with this section until it has complied with the provisions of this section. These fines shall be deposited into the Hospital Building Fund as specified pursuant to Section 129795. A hospital assessed a fine pursuant to this section may appeal the assessment to the Hospital Building Safety Board.
3939
4040 The people of the State of California do enact as follows:
4141
4242 ## The people of the State of California do enact as follows:
4343
4444 SECTION 1. Section 130061 of the Health and Safety Code is amended to read:130061. (a) An owner of a general acute care hospital building that is classified as a nonconforming Structural Performance Category-1 (SPC-1) building, who has not requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than April 15, 2007, describing the status of each building in complying with the requirements of Section 130060. The report shall identify at least all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of this action.(b) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, who has requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than June 30, 2009, describing the status of each building in complying with the requirements of Section 130060. The report shall identify, at a minimum, all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of that action.(c) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, shall submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the requirements of Section 130060, and annually thereafter shall update the office with any changes or adjustments. The report shall identify at least all of the following:(1) For each building that is subject to subdivision (a) of Section 130060 that is planned for retrofit or replacement, the report shall identify:(A) Whether the hospital owner intends to retrofit or replace the building to SPC-2, SPC-3, SPC-4, or SPC-5.(B) The deadline, as described in Section 130060 or 130061.5, for retrofit or replacement of the building that the hospital owner intends to meet, and the applicable extension for which the hospital owner has been approved.(C) The project number or numbers for retrofit or replacement of each building.(D) The projected construction start date or dates and projected construction completion date or dates.(E) The most recent project status and approvals.(F) The number of inpatient beds and patient days, by type of unit and type of service to be provided.(2) For the building or buildings to be removed from acute care service, the following information shall be included:(A) The projected date or dates the building will be removed from service.(B) The planned uses of the building or buildings to be removed from acute care service.(C) The inpatient services currently delivered in the building or buildings.(D) The number of inpatient beds and patient days, by type of unit and type of service, for the years 2008, 2009, and 2010. each of the years 2008 to 2016, inclusive.(E) Whether the general acute care services and beds will be relocated to a new or retrofitted building and any corresponding building sites or project numbers.(3) Each hospital owner shall also report, for each facility for which any buildings will be removed from acute care service, any net change in the number of inpatient beds, by type of unit and type of service, taking into account beds provided in buildings to be taken out of service, beds provided in buildings to be retrofitted or replaced, and beds provided in any other buildings used for general acute care inpatient services by the facility.(4) Each hospital owner shall report any general acute care hospital inpatient service that is provided in any general acute care hospital building that is rated SPC-1.(5) Each hospital owner shall report the final configuration of all buildings on its campus showing how each building will comply with the SPC-5/NPC-4 or 5 requirements, whether by retrofit or by replacement, and the type of services that will be provided in each general acute care hospital building.(d) The office shall make the information required by subdivisions (a) and (b), available on its Internet Web site within 90 days of receipt of this information.(e) The office shall make the information required by subdivision (c) available on its Internet Web site within 90 days of receipt of this information. The office shall include the hospital name, hospital owners, and location of the buildings included in the report, and, to the extent possible, for service areas containing buildings for which hospital owners report information pursuant to subdivision (c), include information on the number of inpatient beds, by type of unit and type of service, provided by facilities operating buildings that are classified as SPC-2, SPC-3, SPC-4, and SPC-5.(f) Hospitals that have not reported pursuant to this section are not eligible for the extension provided in subdivision (f) of Section 130060.(g) A hospital that has not submitted a report pursuant to this section shall be assessed a fine of ten dollars ($10) per licensed acute care bed per day, but in no case to exceed one thousand dollars ($1,000) per day for each SPC-1 building not in compliance with this section until it has complied with the provisions of this section. These fines shall be deposited into the Hospital Building Fund as specified pursuant to Section 129795. A hospital assessed a fine pursuant to this section may appeal the assessment to the Hospital Building Safety Board.
4545
4646 SECTION 1. Section 130061 of the Health and Safety Code is amended to read:
4747
4848 ### SECTION 1.
4949
5050 130061. (a) An owner of a general acute care hospital building that is classified as a nonconforming Structural Performance Category-1 (SPC-1) building, who has not requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than April 15, 2007, describing the status of each building in complying with the requirements of Section 130060. The report shall identify at least all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of this action.(b) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, who has requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than June 30, 2009, describing the status of each building in complying with the requirements of Section 130060. The report shall identify, at a minimum, all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of that action.(c) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, shall submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the requirements of Section 130060, and annually thereafter shall update the office with any changes or adjustments. The report shall identify at least all of the following:(1) For each building that is subject to subdivision (a) of Section 130060 that is planned for retrofit or replacement, the report shall identify:(A) Whether the hospital owner intends to retrofit or replace the building to SPC-2, SPC-3, SPC-4, or SPC-5.(B) The deadline, as described in Section 130060 or 130061.5, for retrofit or replacement of the building that the hospital owner intends to meet, and the applicable extension for which the hospital owner has been approved.(C) The project number or numbers for retrofit or replacement of each building.(D) The projected construction start date or dates and projected construction completion date or dates.(E) The most recent project status and approvals.(F) The number of inpatient beds and patient days, by type of unit and type of service to be provided.(2) For the building or buildings to be removed from acute care service, the following information shall be included:(A) The projected date or dates the building will be removed from service.(B) The planned uses of the building or buildings to be removed from acute care service.(C) The inpatient services currently delivered in the building or buildings.(D) The number of inpatient beds and patient days, by type of unit and type of service, for the years 2008, 2009, and 2010. each of the years 2008 to 2016, inclusive.(E) Whether the general acute care services and beds will be relocated to a new or retrofitted building and any corresponding building sites or project numbers.(3) Each hospital owner shall also report, for each facility for which any buildings will be removed from acute care service, any net change in the number of inpatient beds, by type of unit and type of service, taking into account beds provided in buildings to be taken out of service, beds provided in buildings to be retrofitted or replaced, and beds provided in any other buildings used for general acute care inpatient services by the facility.(4) Each hospital owner shall report any general acute care hospital inpatient service that is provided in any general acute care hospital building that is rated SPC-1.(5) Each hospital owner shall report the final configuration of all buildings on its campus showing how each building will comply with the SPC-5/NPC-4 or 5 requirements, whether by retrofit or by replacement, and the type of services that will be provided in each general acute care hospital building.(d) The office shall make the information required by subdivisions (a) and (b), available on its Internet Web site within 90 days of receipt of this information.(e) The office shall make the information required by subdivision (c) available on its Internet Web site within 90 days of receipt of this information. The office shall include the hospital name, hospital owners, and location of the buildings included in the report, and, to the extent possible, for service areas containing buildings for which hospital owners report information pursuant to subdivision (c), include information on the number of inpatient beds, by type of unit and type of service, provided by facilities operating buildings that are classified as SPC-2, SPC-3, SPC-4, and SPC-5.(f) Hospitals that have not reported pursuant to this section are not eligible for the extension provided in subdivision (f) of Section 130060.(g) A hospital that has not submitted a report pursuant to this section shall be assessed a fine of ten dollars ($10) per licensed acute care bed per day, but in no case to exceed one thousand dollars ($1,000) per day for each SPC-1 building not in compliance with this section until it has complied with the provisions of this section. These fines shall be deposited into the Hospital Building Fund as specified pursuant to Section 129795. A hospital assessed a fine pursuant to this section may appeal the assessment to the Hospital Building Safety Board.
5151
5252 130061. (a) An owner of a general acute care hospital building that is classified as a nonconforming Structural Performance Category-1 (SPC-1) building, who has not requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than April 15, 2007, describing the status of each building in complying with the requirements of Section 130060. The report shall identify at least all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of this action.(b) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, who has requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than June 30, 2009, describing the status of each building in complying with the requirements of Section 130060. The report shall identify, at a minimum, all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of that action.(c) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, shall submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the requirements of Section 130060, and annually thereafter shall update the office with any changes or adjustments. The report shall identify at least all of the following:(1) For each building that is subject to subdivision (a) of Section 130060 that is planned for retrofit or replacement, the report shall identify:(A) Whether the hospital owner intends to retrofit or replace the building to SPC-2, SPC-3, SPC-4, or SPC-5.(B) The deadline, as described in Section 130060 or 130061.5, for retrofit or replacement of the building that the hospital owner intends to meet, and the applicable extension for which the hospital owner has been approved.(C) The project number or numbers for retrofit or replacement of each building.(D) The projected construction start date or dates and projected construction completion date or dates.(E) The most recent project status and approvals.(F) The number of inpatient beds and patient days, by type of unit and type of service to be provided.(2) For the building or buildings to be removed from acute care service, the following information shall be included:(A) The projected date or dates the building will be removed from service.(B) The planned uses of the building or buildings to be removed from acute care service.(C) The inpatient services currently delivered in the building or buildings.(D) The number of inpatient beds and patient days, by type of unit and type of service, for the years 2008, 2009, and 2010. each of the years 2008 to 2016, inclusive.(E) Whether the general acute care services and beds will be relocated to a new or retrofitted building and any corresponding building sites or project numbers.(3) Each hospital owner shall also report, for each facility for which any buildings will be removed from acute care service, any net change in the number of inpatient beds, by type of unit and type of service, taking into account beds provided in buildings to be taken out of service, beds provided in buildings to be retrofitted or replaced, and beds provided in any other buildings used for general acute care inpatient services by the facility.(4) Each hospital owner shall report any general acute care hospital inpatient service that is provided in any general acute care hospital building that is rated SPC-1.(5) Each hospital owner shall report the final configuration of all buildings on its campus showing how each building will comply with the SPC-5/NPC-4 or 5 requirements, whether by retrofit or by replacement, and the type of services that will be provided in each general acute care hospital building.(d) The office shall make the information required by subdivisions (a) and (b), available on its Internet Web site within 90 days of receipt of this information.(e) The office shall make the information required by subdivision (c) available on its Internet Web site within 90 days of receipt of this information. The office shall include the hospital name, hospital owners, and location of the buildings included in the report, and, to the extent possible, for service areas containing buildings for which hospital owners report information pursuant to subdivision (c), include information on the number of inpatient beds, by type of unit and type of service, provided by facilities operating buildings that are classified as SPC-2, SPC-3, SPC-4, and SPC-5.(f) Hospitals that have not reported pursuant to this section are not eligible for the extension provided in subdivision (f) of Section 130060.(g) A hospital that has not submitted a report pursuant to this section shall be assessed a fine of ten dollars ($10) per licensed acute care bed per day, but in no case to exceed one thousand dollars ($1,000) per day for each SPC-1 building not in compliance with this section until it has complied with the provisions of this section. These fines shall be deposited into the Hospital Building Fund as specified pursuant to Section 129795. A hospital assessed a fine pursuant to this section may appeal the assessment to the Hospital Building Safety Board.
5353
5454 130061. (a) An owner of a general acute care hospital building that is classified as a nonconforming Structural Performance Category-1 (SPC-1) building, who has not requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than April 15, 2007, describing the status of each building in complying with the requirements of Section 130060. The report shall identify at least all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of this action.(b) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, who has requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than June 30, 2009, describing the status of each building in complying with the requirements of Section 130060. The report shall identify, at a minimum, all of the following:(1) Each building that is subject to subdivision (a) of Section 130060.(2) The project number or numbers for retrofit or replacement of each building.(3) The projected construction start date or dates and projected construction completion date or dates.(4) The building or buildings to be removed from acute care service and the projected date or dates of that action.(c) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, shall submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the requirements of Section 130060, and annually thereafter shall update the office with any changes or adjustments. The report shall identify at least all of the following:(1) For each building that is subject to subdivision (a) of Section 130060 that is planned for retrofit or replacement, the report shall identify:(A) Whether the hospital owner intends to retrofit or replace the building to SPC-2, SPC-3, SPC-4, or SPC-5.(B) The deadline, as described in Section 130060 or 130061.5, for retrofit or replacement of the building that the hospital owner intends to meet, and the applicable extension for which the hospital owner has been approved.(C) The project number or numbers for retrofit or replacement of each building.(D) The projected construction start date or dates and projected construction completion date or dates.(E) The most recent project status and approvals.(F) The number of inpatient beds and patient days, by type of unit and type of service to be provided.(2) For the building or buildings to be removed from acute care service, the following information shall be included:(A) The projected date or dates the building will be removed from service.(B) The planned uses of the building or buildings to be removed from acute care service.(C) The inpatient services currently delivered in the building or buildings.(D) The number of inpatient beds and patient days, by type of unit and type of service, for the years 2008, 2009, and 2010. each of the years 2008 to 2016, inclusive.(E) Whether the general acute care services and beds will be relocated to a new or retrofitted building and any corresponding building sites or project numbers.(3) Each hospital owner shall also report, for each facility for which any buildings will be removed from acute care service, any net change in the number of inpatient beds, by type of unit and type of service, taking into account beds provided in buildings to be taken out of service, beds provided in buildings to be retrofitted or replaced, and beds provided in any other buildings used for general acute care inpatient services by the facility.(4) Each hospital owner shall report any general acute care hospital inpatient service that is provided in any general acute care hospital building that is rated SPC-1.(5) Each hospital owner shall report the final configuration of all buildings on its campus showing how each building will comply with the SPC-5/NPC-4 or 5 requirements, whether by retrofit or by replacement, and the type of services that will be provided in each general acute care hospital building.(d) The office shall make the information required by subdivisions (a) and (b), available on its Internet Web site within 90 days of receipt of this information.(e) The office shall make the information required by subdivision (c) available on its Internet Web site within 90 days of receipt of this information. The office shall include the hospital name, hospital owners, and location of the buildings included in the report, and, to the extent possible, for service areas containing buildings for which hospital owners report information pursuant to subdivision (c), include information on the number of inpatient beds, by type of unit and type of service, provided by facilities operating buildings that are classified as SPC-2, SPC-3, SPC-4, and SPC-5.(f) Hospitals that have not reported pursuant to this section are not eligible for the extension provided in subdivision (f) of Section 130060.(g) A hospital that has not submitted a report pursuant to this section shall be assessed a fine of ten dollars ($10) per licensed acute care bed per day, but in no case to exceed one thousand dollars ($1,000) per day for each SPC-1 building not in compliance with this section until it has complied with the provisions of this section. These fines shall be deposited into the Hospital Building Fund as specified pursuant to Section 129795. A hospital assessed a fine pursuant to this section may appeal the assessment to the Hospital Building Safety Board.
5555
5656
5757
5858 130061. (a) An owner of a general acute care hospital building that is classified as a nonconforming Structural Performance Category-1 (SPC-1) building, who has not requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than April 15, 2007, describing the status of each building in complying with the requirements of Section 130060. The report shall identify at least all of the following:
5959
6060 (1) Each building that is subject to subdivision (a) of Section 130060.
6161
6262 (2) The project number or numbers for retrofit or replacement of each building.
6363
6464 (3) The projected construction start date or dates and projected construction completion date or dates.
6565
6666 (4) The building or buildings to be removed from acute care service and the projected date or dates of this action.
6767
6868 (b) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, who has requested an extension of the deadline described in subdivision (a) or (b) of Section 130060, shall submit a report to the office no later than June 30, 2009, describing the status of each building in complying with the requirements of Section 130060. The report shall identify, at a minimum, all of the following:
6969
7070 (1) Each building that is subject to subdivision (a) of Section 130060.
7171
7272 (2) The project number or numbers for retrofit or replacement of each building.
7373
7474 (3) The projected construction start date or dates and projected construction completion date or dates.
7575
7676 (4) The building or buildings to be removed from acute care service and the projected date or dates of that action.
7777
7878 (c) An owner of a general acute care hospital building that is classified as a nonconforming, Structural Performance Category-1 (SPC-1) building, shall submit a report to the office no later than November 1, 2010, describing the status of each building in complying with the requirements of Section 130060, and annually thereafter shall update the office with any changes or adjustments. The report shall identify at least all of the following:
7979
8080 (1) For each building that is subject to subdivision (a) of Section 130060 that is planned for retrofit or replacement, the report shall identify:
8181
8282 (A) Whether the hospital owner intends to retrofit or replace the building to SPC-2, SPC-3, SPC-4, or SPC-5.
8383
8484 (B) The deadline, as described in Section 130060 or 130061.5, for retrofit or replacement of the building that the hospital owner intends to meet, and the applicable extension for which the hospital owner has been approved.
8585
8686 (C) The project number or numbers for retrofit or replacement of each building.
8787
8888 (D) The projected construction start date or dates and projected construction completion date or dates.
8989
9090 (E) The most recent project status and approvals.
9191
9292 (F) The number of inpatient beds and patient days, by type of unit and type of service to be provided.
9393
9494 (2) For the building or buildings to be removed from acute care service, the following information shall be included:
9595
9696 (A) The projected date or dates the building will be removed from service.
9797
9898 (B) The planned uses of the building or buildings to be removed from acute care service.
9999
100100 (C) The inpatient services currently delivered in the building or buildings.
101101
102102 (D) The number of inpatient beds and patient days, by type of unit and type of service, for the years 2008, 2009, and 2010. each of the years 2008 to 2016, inclusive.
103103
104104 (E) Whether the general acute care services and beds will be relocated to a new or retrofitted building and any corresponding building sites or project numbers.
105105
106106 (3) Each hospital owner shall also report, for each facility for which any buildings will be removed from acute care service, any net change in the number of inpatient beds, by type of unit and type of service, taking into account beds provided in buildings to be taken out of service, beds provided in buildings to be retrofitted or replaced, and beds provided in any other buildings used for general acute care inpatient services by the facility.
107107
108108 (4) Each hospital owner shall report any general acute care hospital inpatient service that is provided in any general acute care hospital building that is rated SPC-1.
109109
110110 (5) Each hospital owner shall report the final configuration of all buildings on its campus showing how each building will comply with the SPC-5/NPC-4 or 5 requirements, whether by retrofit or by replacement, and the type of services that will be provided in each general acute care hospital building.
111111
112112 (d) The office shall make the information required by subdivisions (a) and (b), available on its Internet Web site within 90 days of receipt of this information.
113113
114114 (e) The office shall make the information required by subdivision (c) available on its Internet Web site within 90 days of receipt of this information. The office shall include the hospital name, hospital owners, and location of the buildings included in the report, and, to the extent possible, for service areas containing buildings for which hospital owners report information pursuant to subdivision (c), include information on the number of inpatient beds, by type of unit and type of service, provided by facilities operating buildings that are classified as SPC-2, SPC-3, SPC-4, and SPC-5.
115115
116116 (f) Hospitals that have not reported pursuant to this section are not eligible for the extension provided in subdivision (f) of Section 130060.
117117
118118 (g) A hospital that has not submitted a report pursuant to this section shall be assessed a fine of ten dollars ($10) per licensed acute care bed per day, but in no case to exceed one thousand dollars ($1,000) per day for each SPC-1 building not in compliance with this section until it has complied with the provisions of this section. These fines shall be deposited into the Hospital Building Fund as specified pursuant to Section 129795. A hospital assessed a fine pursuant to this section may appeal the assessment to the Hospital Building Safety Board.